Temperature/Pressure Data Logger EBI-125 A-PT with
flexible probes |
|
Technical
Data
|
Type |
Pressure
|
Temperature |
Memory
|
6.000 readings
|
2 x 6.000
readings
|
Measuring range
|
0 ... 4 bar (Burst pressure
10 bar) |
-40 ...
+150°C (+150°C up to max. 1 h) |
Accuracy
|
±0,5% of measuring
range
|
±0.1°C ± 1 Digit |
Resolution
|
1 mbar |
0.1°C |
Measuring probe
|
ceramic pressure
sensor
|
2 x Pt 1000, Ø 1,5
mm |
Probe
length |
|
2 flexible probes, 500 mm each, diameter 1,5
mm |
Pressure resistance of
housing
|
|
0 bar ... 10 bar |
Measuring rate
|
|
selectable from
1s ... 8 h |
Measuring mode
|
|
endless, Start/Stop,
Start with predetermined measuring rate
|
Battery life
|
|
approx. 1-3 years
(depending on measuring rate) |
Dimensions
|
|
Ø 48 mm x 28 mm |
Weight
|
|
200 g |
Protection class
|
|
IP 68 |
Certificate
|
|
9 points according to
DIN ISO 9001 |
|
Applications |
- Monitors temperature and pressure in chemical/pharmaceutical
industry
- Monitors temperature and pressure in autoclaves
- Controls pressure in heating systems
|
Attributes |
- Programming and evaluation with PC
- Waterproof
- High accuracy
- According to EN 285/554, DIN 58946-6 and DIN EN 13060
- Factory calibration
certificate
| |